1. The 7-inch touch screen is industrial-grade and displays detailed information.
2.All-new metal furnace body structure with improved baseline and precision.
The heating uses indirect conduction, which offers great uniformity and stability, lowers pulse radiation, and outperforms standard heating modes.
3. USB communication interface: high adaptability, dependable and uninterrupted communication, and self-recovery connection function.
Follow us on Facebook and Youtube.
Click this link to find out more about thermal analyzers.
SKZ1052 Differential Scanning Calorimeter (DSC)
Performance testing, pleasant control, and polymer lookup all make full-size use of DSC.The purpose is to decide how temperature and warmth glide relate to the internal warmth transition.DSC lookup and improvement encompasses the domains of glass transition temperature,melting point, segment transition, bloodless crystallization, and oxidation induction time (OIT).
Features:
1. The 7-inch touch screen is industrial-grade and displays detailed information.
2.All-new metal furnace body structure with improved baseline and precision.
The heating uses indirect conduction, which offers great uniformity and stability, lowers pulse radiation, and outperforms standard heating modes.
3. USB communication interface: high adaptability, dependable and uninterrupted communication, and self-recovery connection function.
4. Automatically switch between two channels of atmospheric flow, with a fast switching speed and a short stabilizing time. At the same time, a protective gas input is introduced.
5. The program is straightforward and easy to use.
Standards:
ISO/TR10837:1991, ASTM D3895-1998, ASTM E 967, ASTM E 968, ASTM E 793, ASTM D 3895, ASTM D 3417, ASTM D 3418, ISO 11357-6
Temperature range | RT~ 550 ℃ |
Heating rate | 0.1~80℃/min |
Temperature resolution | 0.001℃ |
Temperature fluctuations | ±0.001℃ |
Temperature repeatability | ±0.1℃ |
Constant temperature time | Program setting ≤24h |
DSC range | 0~±600Mw |
DSC noise | 0.01Μw |
DSC resolution | 0.01Μw |
DSC accuracy | 0.01Mw |
Atmosphere control gas | Nitrogen, Oxygen (instrument switch automatically) |
Gas flow rate | 0-300Ml/min |
Gas pressure | 0.2Mpa |
Control Mode | Rising temperature, constant temperature (fully automatic) |
Curve scanning | Rising scan |
Gas flow control | Built-in digital flow meter & Software control, automatic |
Display | 24-bit 7 inches LCD touch screen display |
Data interface | standard USB connector |
Parameter standard | With standard material, with calibration function, the user may correct temperature and heat enthalpy |
Software | operating software |
Thermocouples | Total 3 sets of thermocouples for: 1). Test sample temperature 2). test furnace temperature, 3). Test internal ambient temperature |
Cooling device | Air cool |
Power | AC220V/50Hz |